smartline temperature transmitters
SmartLine STT650
STT650 DIN rail mounted temperature transmitters offer high measurement accuracy, stability and reliability over a wide range of process and ambient temperatures.
Please sign in to access more documents
Once signed in, you may be able to access additional documents for your account.
The SmartLine STT650 temperature transmitters offer the following features:
Leading Performance
- High accuracy – 0.1°C for RTD
- Faster response, up to 135 ms update
- Stable measurement, 0.12% of span for two years
- Built-in 1500 VAC galvanic isolation
- Superior noise performance – tested for 2 KV surge/2.5 kv burst noise
- Sensor diagnostics
- Simple and easier configuration
- Compact design, saving over 50% panel space.
Standards & Certifications
- Approvals
- CE Mark, IECEx, ATEX, CSA, FM Approvals (Intrinsic Safety, Non-Incendive, Dust Tight)
I/O Capabilities
- Update/Response Time
- Analog Output: 135 ms, HART Output: 440 ms
- Sensor Error Detection
- NAMUR NE43 Compliant
- Galvanic Isolation
- 1500 VAC
- Configuration
- STT17C PC Configuration Tool (Windows XP/7/8/8.1, graphical UI, non-hazardous environments)
Electrical Specifications
- Supply Voltage
- 8 to 30 VDC (Intrinsically Safe)
- 8 to 35 VDC (Non-Intrinsically Safe)
- Power Supply Effect
- ≤ 0.005% of span per VDC
- Output Options
- 4 to 20 mA Analog, HART® (v7.0)
- Input Type
- Universal Input: RTD (2, 3, 4 wire), Thermocouple (B, E, J, K, L, N, R, S, T, U, W3, W5, +Lr), Ohm, mV voltage direct current
Environmental Conditions
- Stability
- ±0.12% of span Over 2 years
- Relative Humidity
- 0 to 95% RH (Non-Condensing)
- Operating Temperature Range
- -40 to 85 deg C (-40 to 185 deg F)
- Load Resistance
- Analog: ≤ (V - 8)/0.023 A, HART: ≤ (V - 7.2)/0.023 A
Applications & Compatibility
- Mounting
- DIN Rail, Compact Design
Standards & Certifications
- Approvals
- CE Mark, IECEx, ATEX, CSA, FM Approvals (Intrinsic Safety, Non-Incendive, Dust Tight)
I/O Capabilities
- Update/Response Time
- Analog Output: 135 ms, HART Output: 440 ms
- Sensor Error Detection
- NAMUR NE43 Compliant
- Galvanic Isolation
- 1500 VAC
- Configuration
- STT17C PC Configuration Tool (Windows XP/7/8/8.1, graphical UI, non-hazardous environments)
Electrical Specifications
- Supply Voltage
- 8 to 30 VDC (Intrinsically Safe)
- 8 to 35 VDC (Non-Intrinsically Safe)
- Power Supply Effect
- ≤ 0.005% of span per VDC
- Output Options
- 4 to 20 mA Analog, HART® (v7.0)
- Input Type
- Universal Input: RTD (2, 3, 4 wire), Thermocouple (B, E, J, K, L, N, R, S, T, U, W3, W5, +Lr), Ohm, mV voltage direct current
Environmental Conditions
- Stability
- ±0.12% of span Over 2 years
- Relative Humidity
- 0 to 95% RH (Non-Condensing)
- Operating Temperature Range
- -40 to 85 deg C (-40 to 185 deg F)
- Load Resistance
- Analog: ≤ (V - 8)/0.023 A, HART: ≤ (V - 7.2)/0.023 A
Applications & Compatibility
- Mounting
- DIN Rail, Compact Design
- Approvals : CE Mark, IECEx, ATEX, CSA, FM Approvals (Intrinsic Safety, Non-Incendive, Dust Tight)
- Update/Response Time : Analog Output: 135 ms, HART Output: 440 ms
- Supply Voltage : 8 to 30 VDC (Intrinsically Safe)|8 to 35 VDC (Non-Intrinsically Safe)
- Stability : ±0.12% of span Over 2 years
- Sensor Error Detection : NAMUR NE43 Compliant
- Relative Humidity : 0 to 95% RH (Non-Condensing)
- Power Supply Effect : ≤ 0.005% of span per VDC
- Output Options : 4 to 20 mA Analog, HART® (v7.0)
- Operating Temperature Range : -40 to 85 deg C (-40 to 185 deg F)
- Mounting : DIN Rail, Compact Design
- Load Resistance : Analog: ≤ (V - 8)/0.023 A, HART: ≤ (V - 7.2)/0.023 A
- Input Type : Universal Input: RTD (2, 3, 4 wire), Thermocouple (B, E, J, K, L, N, R, S, T, U, W3, W5, +Lr), Ohm, mV voltage direct current
- Galvanic Isolation : 1500 VAC
- Configuration : STT17C PC Configuration Tool (Windows XP/7/8/8.1, graphical UI, non-hazardous environments)
Brochure
Data Sheet
Filters
Please sign in to view part numbers available for purchase based on your account Sign In