Conductivity
4973 Flow Thru Mounting
Honeywell’s 4973 series of conductivity cells ensures reliable and accurate conductivity measurement, and is suitable for all-purpose applications.
Please sign in to access more documents
Once signed in, you may be able to access additional documents for your account.
What Is It?
The 4973 series ensures reliable, continuous measurement of electrolytic conductivity in industrial water processes where the temperature and pressure limits are high.
Made of polyethersulfone (PES) for high-corrosion resistance, the cells are available with titanium electrodes (for 0.01 and 0.1 cell constants) and high-density graphite (for 1.0 and 10.0 cell constants). The wide array of cell constants available makes them ideal for diverse applications.
With multiple mounting options such as, flow-thru and in-line, the 4973 series is easy to install. The 4974 style uses this series of cells for sanitary installations. The cells are mounted using an industry standard tri-clamp flange, a common mounting arrangement.
How Does It Work?
Conductivity cells actually measure the resistance of the liquid they are in. The 4973 series provides continuous, reliable measurements at temperatures up to 140°C and pressures up to 250 psig.
For in-line applications these cells can be installed directly into the process pipe in a separately supplied ¾-inch tee or in a by-pass stream. For flow-thru applications the cell is mounted in a separately supplied flow chamber.
The cells keep the electrodes and the temperature compensator immersed in the stream flow, ensuring that the cell responds quickly and accurately to changes in both solution concentration and temperature. For sanitary clean-in-place (CIP) piping systems, the 4974 style includes standard 1 ½-inch or 2-inch CIP fittings.
What Problems Does It Solve?
The 4973 series simplifies the conductivity measurement process and improves overall process efficiency. It provides users with a reliable, cost-effective and accurate solution and is suited for a wide range of applications, including high purity water measurements.
The rugged design of this series ensures easy installation, maintenance and replacement.
Applications & Compatibility
- Applications
- Reliable, continuous measurement of electrolytic conductivity in industrial water processes
Environmental Conditions
- Temperature Rating
- Up to 140 deg C (284 deg F)
- Pressure Ratings
- Up to 250 psig (1724 kPa)
I/O Capabilities
- Mounting Types
- Flow-Through (Direct in process stream or bypass sample line with 3/4 in pipe tee)
- Insertion Depth
- 64 mm (2.5 in) for 0.1 Constant
- 89 mm (3.5 in) for 1, 10, and 0.01 Constants
Mechanical & Physical Properties
- Electrode Materials
- Titanium (0.01 and 0.1), High-Density Graphite (1 and 10)
Technical Specifications
- Electrical Connections
- (4) Leads with integral Automatic Temperature Compensator
- Quick disconnect option (Turck)
Communication & Protocols
- Cell Constants Supported
- 0.01, 0.1, 1, and 10 cm⁻¹
Material Specifications
- Cell Body Material
- Polyethersulfone (PES) for high corrosion resistance
Network & Connectivity Infrastructure
- Cable/Leadwire
- Integral PVC-covered, non-shielded, 18-gage cable
- Quick disconnect option available
Standards & Certifications
- Approvals
- ASME Section VIII, Div. 1, UG-101, CRN #0F11607-5C
Applications & Compatibility
- Applications
- Reliable, continuous measurement of electrolytic conductivity in industrial water processes
Environmental Conditions
- Temperature Rating
- Up to 140 deg C (284 deg F)
- Pressure Ratings
- Up to 250 psig (1724 kPa)
I/O Capabilities
- Mounting Types
- Flow-Through (Direct in process stream or bypass sample line with 3/4 in pipe tee)
- Insertion Depth
- 64 mm (2.5 in) for 0.1 Constant
- 89 mm (3.5 in) for 1, 10, and 0.01 Constants
Mechanical & Physical Properties
- Electrode Materials
- Titanium (0.01 and 0.1), High-Density Graphite (1 and 10)
Technical Specifications
- Electrical Connections
- (4) Leads with integral Automatic Temperature Compensator
- Quick disconnect option (Turck)
Communication & Protocols
- Cell Constants Supported
- 0.01, 0.1, 1, and 10 cm⁻¹
Material Specifications
- Cell Body Material
- Polyethersulfone (PES) for high corrosion resistance
Network & Connectivity Infrastructure
- Cable/Leadwire
- Integral PVC-covered, non-shielded, 18-gage cable
- Quick disconnect option available
Standards & Certifications
- Approvals
- ASME Section VIII, Div. 1, UG-101, CRN #0F11607-5C
- Applications : Reliable, continuous measurement of electrolytic conductivity in industrial water processes
- Temperature Rating : Up to 140 deg C (284 deg F)
- Pressure Ratings : Up to 250 psig (1724 kPa)
- Mounting Types : Flow-Through (Direct in process stream or bypass sample line with 3/4 in pipe tee)
- Insertion Depth : 64 mm (2.5 in) for 0.1 Constant|89 mm (3.5 in) for 1, 10, and 0.01 Constants
- Electrode Materials : Titanium (0.01 and 0.1), High-Density Graphite (1 and 10)
- Electrical Connections : (4) Leads with integral Automatic Temperature Compensator|Quick disconnect option (Turck)
- Cell Constants Supported : 0.01, 0.1, 1, and 10 cm⁻¹
- Cell Body Material : Polyethersulfone (PES) for high corrosion resistance
- Cable/Leadwire : Integral PVC-covered, non-shielded, 18-gage cable|Quick disconnect option available
- Approvals : ASME Section VIII, Div. 1, UG-101, CRN #0F11607-5C
Application Note
Brochure
Manuals and Guides